But recently, Valley, an assistant professor in the University of Michigan’s Division of Pulmonary and Critical Care, realized first-hand that the small device may yield less accurate oxygen readings in patients with dark skin.
One end of the device sends light through the finger while a sensor on the other side receives this light and uses it to detect the color of your blood; bright red blood is highly oxygenated, while blue or purplish blood is less. If the device isn’t calibrated for darker skin tones, the pigmentation of the skin could affect how the light is absorbed by the sensor, leading to flawed oxygen readings.
On a person of color, the pulse oximeter could indicate a patient’s oxygen levels are normal, suggesting they could be discharged home — but then their blood samples might show low oxygen levels, suggesting they not only needed additional care but also oxygen support.
“At the time, it was during the spring 2020 surge in Michigan, and we didn’t actually put together that this was a problem of race,” Valley told CNN. “We thought potentially, this was a problem of Covid, because that’s what we were being inundated with — hundreds of patients with Covid.”
Before the pandemic, the majority of Valley’s patients at Michigan Medicine were White. But as the hospital treated more Covid-19 patients, Valley realized that many were Black or brown patients coming from other overburdened facilities.
“At 92%, we typically don’t make clinical changes based on that, but if that 92% on a pulse oximeter means that their oxygen levels are actually below 88%, well, that is something that I would make a clinical change about — whether it’s starting someone on oxygen, or whether it’s increasing their oxygen values,” Valley said.
“I still struggle with what to do when I have a Black patient who has a pulse oximeter value that’s marginal,” he said. “I think for an individual at home, I think it heightens the need to take your symptoms into account.”
‘The consequences of the neglect … has been revealed’
But only just recently the public has been made more aware of this health disparity, and US health officials have announced plans to investigate the accuracy of pulse oximeters.
“It’s problematic,” Valley said.
“There was a study just like ours done in 1990, but at a smaller scale at a single center. The problem from 1990 to 2020 is really a problem in dissemination and education. In medical school, in residency, in fellowship, I didn’t know about this problem,” he said. “It’s not that we’re now using pulse oximeters more often. I use a pulse oximeter every day, multiple times a day, all day in the intensive care unit, and that’s unchanged from 10 years ago. It’s almost like my right hand in an intensive care unit. We depend on what someone’s oxygen levels are.”
Some experts view the pandemic as shedding more of a spotlight on the pulse oximeter’s limitations in measuring oxygen levels among Black and brown patients.
“People have known that there’s an issue with the over-estimation of these oxygen saturation levels, particularly in dark-skinned patients, but they did not appreciate the significance of this issue until the Covid-19 pandemic,” she said. “The pandemic has seen an increased use of pulse oximeters in the hospital and in the home setting and thus the consequences of the neglect of the issues with errors in these pulse oximeter has been revealed.”
Jakachira and her research adviser Kimani Toussaint argue that in the years following Aoyagi’s invention, not enough clinical testing was done to improve the device to ensure equity across all demographics and skin tones.
“So, one thing that should be done immediately is improved calibration testing,” Jakachira said, adding that currently, guidelines from the US Food and Drug Administration recommend such testing on pulse oximeters include at least 10 healthy subjects that vary in age and gender with a range of skin pigmentations, of which two individuals or 15% of the group — “whichever is larger,” the FDA guidance indicates — have dark skin.
“This is probably not sufficient and not to draw statistical conclusions on disparities,” she said. “So, using a more diverse population, calibration would be a great first step.”
Future of pulse oximetry
“What we’re working on, we’re trying to mitigate the skin tone issues by doing something interesting with the light, but it’s a significant challenge and this really highlights the need to have diversity and inclusion,” said Toussaint, a professor and senior associate dean in the School of Engineering at Brown University.
When pulse oximeters send beams of light through the fingertips to gauge blood oxygen levels, they are measuring how much oxygen has been absorbed by a certain molecule in the blood called hemoglobin. It turns out that both melanin and hemoglobin absorb light at similar wavelengths and separating out their relative contributions is what can be challenging. Melanin is the substance that produces pigmentation in our skin, hair and eyes.
“So, the melanin will overlap with the absorption properties of the hemoglobin in your blood,” Toussaint said, which can lead to the pulse oximeter’s flawed blood oxygen readings, as different people have different amounts of melanin.
The approach that Jakachira and Toussaint are taking in their quest for a solution is to try to cancel out the effect that absorbed melanin can have on how pulse oximeters measure blood oxygen levels.
“That’s something that we think would be certainly a contribution but maybe translatable to other similar-based technologies that are not just pulse oximeters,” Toussaint said, adding that he could not share additional details of this work as the research team is currently completing a patent application.
FDA advisers to discuss pulse oximeter accuracy
A patient’s oxygen levels are sometimes called the “fifth vital sign,” and there is growing distress around how such investigations into how pulse oximeter inaccuracies among people of color did not occur sooner, according to an editorial that published Monday in the journal JAMA Internal Medicine.
“While the technical limitations of the pulse oximeter have been recognized for more than 30 years, investigations into clinical implications of hidden hypoxemia in patients with darker skin have been relatively (and distressingly) recent,” Dr. Eric Ward of the University of California, San Francisco, and Dr. Mitchell Katz of NYC Health + Hospitals, wrote in their editorial.
They described one plan of action for change.
“Devices exist that perform more equitably but have never been widely distributed,” Ward and Katz wrote. “Health care systems, including academic centers, are large-scale purchasers of pulse oximeters. If they make a commitment to buy only devices that function across skin tones, manufacturers would respond.”
#Scientists #searching #solutions #studies #show #pulse #oximeters #dont #work #people #color